Atlas FDA

Animas IR 1250 Insulin Pump.

FDA device recall Z-0929-05 · posted 2005-06-25 · Terminated

Recall details

Recalling firm
Animas Corporation
Reason for recall
pump failed to respond to keypad button pressess after using the food database feature
Action taken
The recalling firm issued a recall notice dated 4/11/05 to all patients and healthcare providers via first class mail. The letter instructs the patient to reset the pump, thus correcting any problems caused by the software bug, and to discontinue use of the food database. All distributors were contacted by phone on 4/12/05 and instructed to return all pumps in their inventory. On 4/25/05, the recalling began emailing and telephoning patients to arrange for replacement and return of the devices.
Root cause
Other
Status
Terminated
Product code
LZG
Quantity affected
2081 units
Distribution
The products were shipped to distributors in FL, TX, UT, and VA. The products were also shipped to patients nationwide.
Date initiated
2005-04-11
Date posted
2005-06-25
Date terminated
2006-01-24
Location
West Chester, PA

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
ANIMAS MODEL IR1250 INSULIN INFUSION PUMP (K042873)Animas Corp.2004-12-10