Anesthesia device service kits. ASSY-MSN, VENT ENGINE ASSY, 1503-8101-000 ASSY-MSN, VENT ENGINE ASSY 7900, 1009-8216-000
FDA device recall Z-0682-2016 · posted 2016-01-21 · Terminated
Recall details
- Recalling firm
- GE Medical Systems, LLC
- Reason for recall
- The Drive Gas Check Valve of certain Anesthesia products could become stuck in a fixed open position, which could cause pressure to build in the mechanical ventilation cycle. If this issue is left unresolved, it could result in excessive or prolonged pressure in the patient breathing circuit during ventilation potentially resulting in barotrauma.
- Action taken
- GE Healthcare sent an Urgent Medical Device Correction letter dated December 15, 2015, to all affected consignees. The letter described the Safety Issue, Safety Instructions, Affected Product Details, Product Correction and Contact Information. Customers with questions were instructed to contact GE Healthcare Service at 1-800-437-1171 or their local Service Representative.
- Root cause
- Nonconforming Material/Component
- Status
- Terminated
- Product code
- BSZ
- Quantity affected
- 240
- Distribution
- Worldwide Distribution - US (nationwide including PR except NH, VT, & WY. Internationally to ALBANIA, ALGERIA, ARGENTINA, AUSTRALIA, AUSTRIA, BANGLADESH, BELGIUM, BOLIVIA, BOSNIA, BRAZIL, CAMBODIA, CA
- Date initiated
- 2015-12-15
- Date posted
- 2016-01-21
- Date terminated
- 2016-09-21
- Location
- Waukesha, WI