Atlas FDA

Amsco Sonic Energy Console, Part Nos.: B602047470, P080000002, P080000003, B018905015, B018905016, B018905017, B01890501

FDA device recall Z-1408-06 · posted 2006-08-25 · Terminated

Recall details

Recalling firm
Steris Corporation
Reason for recall
Smoke, sparking and fire hazard-Units manufactured between November 2000 and January 2005 are susceptible to water damage at the lid switch cable wiring harness, which may result in smoking, sparking, and fire.
Action taken
The firm sent out a certified letter to all affected consignees alerting them to problems with the device on 6/22/06.
Root cause
Other
Status
Terminated
Product code
FLG
Quantity affected
353 units
Distribution
Worldwide USA and countries of Japan, Canada, Hong Kong, Korea, Lebanon, Qatar, Poland, Russia, Saudi Arabia, Singapore, Taiwan, Germany, Northern Mariana Islands, and Malaysia.
Date initiated
2006-06-22
Date posted
2006-08-25
Date terminated
2007-01-08
Location
Montgomery, AL