Amsco Century Medium Steam Sterilizer (26" x 37.5") (660 x 950mm).
FDA device recall Z-1193-2007 · posted 2007-08-23 · Terminated
Recall details
- Recalling firm
- Steris Corp
- Reason for recall
- The cables inside the pump conduit box may be exposed (not covered with wire nut and/or electrical tape) that could cause a potential electrical shock hazard.
- Action taken
- By telephone 3/22/2007, customers were instructed to cease using the product and a service technician would be dispatched immediately to repair the unit.
- Root cause
- Employee error
- Status
- Terminated
- Product code
- FLE
- Quantity affected
- 78
- Distribution
- Worldwide, including USA, Denmark, Belgium, Greece, India, and Russia.There are no Canadian accounts.
- Date initiated
- 2007-03-22
- Date posted
- 2007-08-23
- Date terminated
- 2010-09-28
- Location
- Mentor, OH