Amplicor [3] AV-HRP. (Avidin-HRP (Horseradish Peroxidase) BGG Conjugate.) Kit: Amplicor IC Detection Kit, 96 test [Gener

FDA device recall Z-1363-04 · posted 2004-08-19 · Terminated

Recall details

Recalling firm
Roche Molecular Systems Inc
Reason for recall
An increased frequency of 'blue foci' that potentially can cause elevated A450 background in microwell plate wells after the addition of conjugate reagent during PCR detection.
Action taken
With respect to U.S distribution, email/letter on 6/24/04 to the local safety officers and general managers stationed at the affiliated distribution center, Roche Diagnostics Corporation of Indianapolis, IN. With respect to overseas distributors, email/letter on 6/24/04 to the local safety officers and general managers stationed at the affiliated distribution centers. Roche stated that the local safety officer is responsible to communicate the advisory both to the customers/end-users and the regulatory authority of the country. On July 6, 2004, a Revised Important Product Advice Notice was sent out to affiliates requiring the immediate discontinuation of the use of the affected kits, and destruction of any kits remaining in customers hands.
Root cause
Other
Status
Terminated
Quantity affected
2482 kits
Distribution
♦ Amplicor CT/NG CT Detection Kit, Lot No.E12231, Material No.20759392018, Expiration 2/2005 , Shipped To: Roche Diagnostics Corporation, 9115 Hague Road, P.O. Box 50457 Indianapolis, IN 46250-0457. ♦
Date initiated
2004-07-07
Date posted
2004-08-19
Date terminated
2005-01-07
Location
Belleville, NJ