Atlas FDA

Amplatz Goose Neck Microsnare Kit, Model SK400 (175 cm length x 4mm loop snare x .018'',0.46mm) and SK701 (200 cm length

FDA device recall Z-0894-06 · posted 2006-06-02 · Terminated

Recall details

Recalling firm
Ev3, Inc
Reason for recall
ev3 has determined that one test unit from each of two lots did not meet mininum tensile strength requirements and therefore the entire lots are being recalled.
Action taken
The recall was initiated on August 16, 2005. All units distributed have been accounted for and the firm completed the recall on December 29, 2005. The firm sent a notification to customers with a Device Recall Field Action Form to record if devices remain at the sight and include with devices returned for credit or replacement.
Root cause
Other
Status
Terminated
Product code
DXE
Quantity affected
58
Distribution
AL, AR, CA, CO, FL, IL, KS, MA, MO, NY, PA, TX, WA, VA, Canada, Germany, Japan
Date initiated
2005-08-16
Date posted
2006-06-02
Date terminated
2006-12-26
Location
Plymouth, MN

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
AMPLATZ GOOSE-NECK SNARE (K901502)Microvena Corp.1990-12-03