Amico E-Cylinder Canister Holder, distributed by Stryker Communications, Flower Mound, TX 75028. Canister holder is used
FDA device recall Z-1569-2009 · posted 2009-07-13 · Terminated
Recall details
- Recalling firm
- Stryker Communications Corp
- Reason for recall
- Design flaw on the E-Cylinder canister holder may result in early fatigue and failure of the bracket screws, which could cause the cage to become detached from its mount and fall.
- Action taken
- An "Urgent: Device Recall" letter dated May 19, 2009 was issued to customers. Customers were advised to discontinue use of all affected E-cylinder holders. Consignees were asked to return the "Product Accountability Form" included with the recall letter, after which consignees will be provided with an improved cylinder holder at no cost. Direct questions to Stryker Communications at 1-972-410-7100.
- Root cause
- Device Design
- Status
- Terminated
- Product code
- BRY
- Quantity affected
- 66 units.
- Distribution
- Nationwide and Canada.
- Date initiated
- 2009-05-19
- Date posted
- 2009-07-13
- Date terminated
- 2010-04-26
- Location
- Flower Mound, TX