Atlas FDA

AMBU / KING LTSD, SIZE 0,1 W/SUCTION PORT, STERILE- Inserted into a patient s pharynx through the mouth to provide paten

FDA device recall Z-0269-2025 · posted 2024-10-30 · Open, Classified

Recall details

Recalling firm
King Systems Corp. dba Ambu, Inc.
Reason for recall
Exceed the limitations to the 510 (k) exemption in 21 CFR 868.9(a) because it introduces a new indication for pediatric use in an Oropharyngeal Airway.
Action taken
King Systems issued Urgent Field Safety Notice on 9/26/624 via email/Certified Mail. Letter states reason for recall, health risk and action to take: You must discontinue use and dispose of King LTS-D" Disposable Laryngeal Tube pediatric sizes and identify alternative devices for appropriate airway management. Within one month of receipt of this notice, please provide return response for this Field Safety Notice. This Notice needs to be passed on to all those who might concern within your organization where the devices could have been transferred. Please transfer this notice to other organizations on which this action has an impact. Please maintain awareness of this notice and resulting action for an appropriate period to ensure effectiveness of the corrective action. Patient safety remains our highest priority. If you have additional questions regarding this information, please contact your local Ambu-King Systems sales representative. Should you have additional questions, please do not hesitate to contact Tammy Feyerherm at tfey@ambu.com.
Root cause
No Marketing Application
Status
Open, Classified
Product code
CAE
Distribution
Nationwide
Date initiated
2024-09-26
Date posted
2024-10-30
Location
Noblesville, IN