ambIT Ambulatory Infusion Pump Cassette; Filter, Spike, Sterile, Vol. 2.3ml, REF 220140, Sorenson Medical, Inc., West Jo
FDA device recall Z-0186-2007 · posted 2006-11-16 · Terminated
Recall details
- Recalling firm
- Sorenson Medical, Inc.
- Reason for recall
- Sterile disposable cassette used with infusion pump may cause an under-infusion when used in a low flow rate therapy of less than 10 ml per hour.
- Action taken
- Consignees were notified by letter on 10/10-11/2006 and told to immediately discontinue use/quarantine all affected cassettes. .
- Root cause
- Other
- Status
- Terminated
- Product code
- FRN
- Quantity affected
- 30,697 units
- Distribution
- Worldwide, including USA, Belgium, France, Germany, Italy, Singapore, Venezuela.
- Date initiated
- 2006-10-10
- Date posted
- 2006-11-16
- Date terminated
- 2011-05-26
- Location
- West Jordan, UT