Atlas FDA

ambIT Ambulatory Infusion Pump Cassette; Filter, Male, Sterile, Vol. 2.3ml, REF 220266, Sorenson Medical, Inc., West Jor

FDA device recall Z-0187-2007 · posted 2006-11-16 · Terminated

Recall details

Recalling firm
Sorenson Medical, Inc.
Reason for recall
Sterile disposable cassette used with infusion pump may cause an under-infusion when used in a low flow rate therapy of less than 10 ml per hour.
Action taken
Consignees were notified by letter on 10/10-11/2006 and told to immediately discontinue use/quarantine all affected cassettes. .
Root cause
Other
Status
Terminated
Product code
FRN
Quantity affected
2000 units
Distribution
Worldwide, including USA, Belgium, France, Germany, Italy, Singapore, Venezuela.
Date initiated
2006-10-10
Date posted
2006-11-16
Date terminated
2011-05-26
Location
West Jordan, UT