Alto MSP (model 617) and Alto 2 MSP (model 627) Implantable cardioverter Defibrillator. Ventricular antitachycardia paci

FDA device recall Z-1598-05 · posted 2005-09-24 · Terminated

Recall details

Recalling firm
ela Medical Llc
Reason for recall
Devices manufactured prior to August 2004 may experience premature battery depletion and/or extended charge times which may affect the device's ability to function.
Action taken
A Dear Doctor letter dated 07/20/05 was sent describing the issue and providing recommendation that physicians follow-up with patients every 3 months. Patients who have frequent episodes of ventricular fibrillation or are pacemaker dependend may require prophylactic explantation or more frequent visits. A list of affected serial numbers was provided to each physician.
Root cause
Other
Status
Terminated
Product code
MRM
Quantity affected
20 total devices (3-model 617 & 17-model 627)
Distribution
Nationwide. AZ, CA, FL, IL, KY, LA, MD, MI, MO, NC, NJ, NM, NY, OH, PA, SC, TN, TX, WA and WI.
Date initiated
2005-07-19
Date posted
2005-09-24
Date terminated
2007-01-28
Location
Plymouth, MN