Alsius CoolGard 3000 Temperature Regulation System

FDA device recall Z-1397-05 · posted 2005-08-24 · Terminated

Recall details

Recalling firm
Alsius Corporation
Reason for recall
Firm wants users to be aware not to leave the units in standby mode for more than an hour in rooms in excess of 75 F due to risk of an alarm mode being generated due to overheating.
Action taken
Firm will be sending letters of caution to all users that reemphasizes labeling beginning 7/25/2005. New software is being developed which will include a change to maintain STANDBY coolant temperature regulation indefinitely, removing any risk of failure.
Root cause
Other
Status
Terminated
Product code
NCX
Quantity affected
207
Distribution
Nationwide and in Uruguay, Netherlands, Switzerland, Spain, Poland, China, Norway, Malasia, Italy, Ireland, greece, Germany, France, Finland, Canada, Belgium, Austria, Bahamas,
Date initiated
2005-07-25
Date posted
2005-08-24
Date terminated
2007-01-31
Location
Irvine, CA