Allura Xper FD20/15; Product Code: 722058; Associated 510(k) numbers: K102005, K130842, K130638, K133292, K141979, K1615
FDA device recall Z-2633-2025 · posted 2025-09-25 · Open, Classified
Recall details
- Recalling firm
- PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
- Reason for recall
- Systems may not perform as intended due to the deterioration of certain internal components (the CMOS battery, hard disk drive, and/or power supply unit). If this deterioration occurs, motorized movements of the system will become unavailable. Manual movements of the stand (longitudinal, transversal and rotational for ceiling mounted systems and rotational for floor mounted systems) and table longitudinal movements remain available. Imaging (X-ray) functionality remains available.
- Action taken
- On September 3, 2025, URGENT MEDICAL DEVICE CORRECTION letters were sent to customers. Actions to be taken: 1. "Circulate this Urgent Medical Device Correction to all users of the system so that they are aware of the issue and follow the instructions included in the letter. 2. In case the affected system has been transferred to another organization, please send a copy of this Urgent Medical Device Correction to that organization and inform Philips about this transfer through your local Philips representative. 3. Keep this Urgent Medical Device Correction with the documentation of the system until Philips corrects your system. Ensure that the letter is in a place likely to be seen/viewed. 4. Complete and return the response form included in this Urgent Medical Device Correction to Philips promptly and no later than 30 days from receipt. 5. If you experience the issue described in this letter, please report it to your local Philips representative. Actions Planned by Recalling Firm: 1. Philips will be replacing the Geo PC in all affected systems. 2. Philips expects to start replacing the Geo PC units by September 2025. Your local Philips representative will contact you to schedule a visit to confirm if your system has a model Wolfdale (Q35 generation) Geo PC and to plan the replacement of the Geo PC once available. 3. If you need additional information or support concerning these issues, contact your local Philips representative at 1-800-722-9377 (Philips Customer Care Solutions Center).
- Root cause
- Device Design
- Status
- Open, Classified
- Product code
- OWB
- Quantity affected
- 98 units (23 US, 75 OUS)
- Distribution
- Worldwide distribution - US Nationwide and the countries of Albania, Algeria, Argentina, Australia, Austria, Azerbaijan, Bangladesh, Belarus, Belgium, Brazil, Bulgaria, Canada, Chile, China, Colombia,
- Date initiated
- 2025-09-03
- Date posted
- 2025-09-25
- Location
- Best, Netherlands
Monitor recalls programmatically
Recalls and clearances share one API, with change alerts when new recalls post.
Related 510(k) clearances
| Record | Firm | Date |
|---|---|---|
| Allura Xper FD series and Allura Xper OR Table series (K161563) | Philips Medical Systems Nederland B.V. | 2016-07-29 |
| ALLURA XPER FD SERIES / ALLURA XPER OR TABLE SERIES (K141979) | Philips Medical Systems Netherland BV | 2014-08-19 |
| ALLURA XPER FD; OR TABLE (K133292) | Philips Medical Systems Nederland B.V. | 2014-03-05 |
| ALLURACLARITY XPER FD SERIES X-RAY SYSTEM (K130638) | Philips Medical Systems Nederland B.V. | 2013-06-28 |
| ALLURA XPER FD OR TABLE SERIES; ALLURA XPER FD10 SERIES; ALLURA XPER FD20 SERIES; ALLURA XPER FD BIPLANE SERIES (K130842) | Philips Medical Systems Nederland B.V. | 2013-05-21 |
| ALLURA XPER OR TABLE SERIES (K102005) | Philips Medical Systems North America, Inc. | 2010-08-09 |