Allura Xper FD20/10; Model Number: 722029;
FDA device recall Z-1076-2026 · posted 2026-01-13 · Open, Classified
Recall details
- Recalling firm
- PHILIPS MEDICAL SYSTEMS NEDERLAND B.V.
- Reason for recall
- Philips has identified that in a limited number of impacted systems the drip tray beneath the cooling unit was not installed under previous field corrections, or evidence of the implementation was not adequately documented. If it is not installed, there is a potential for coolant liquid to contact electrical components. This could lead to electrical short-circuits, which may trigger system fuses and result in system shutdown.
- Action taken
- On December 15, 2025, URGENT Medical Device Correction letters were sent to customers. Actions to be taken: a. Affected systems may continue to be used in accordance with their Instructions for Use (IFU). b. Circulate this letter to all users so that they are aware of the issue. Keep this letter with the documentation of the system until Philips inspects and, where needed, corrects your system. c. Establish an emergency protocol prior to all applicable diagnostics, interventional and minimally invasive procedures to manage the situation should you experience the issue during a procedure. d. If you detect a burning odor, smoke, or fire originating from the R-cabinet in the technical room, immediately switch off the system s main power. Do not operate or use the system further. Follow your institution s emergency procedures. e. In case the affected system has been transferred to another organization, please send a copy of this letter to that organization and inform Philips about this transfer through your local Philips representative. f. Complete and return the attached response form to Philips promptly and no later than 30 days from receipt. g. Should you experience the issue reported in this letter, please report the event to Philips through your local Philips representative.
- Root cause
- Other
- Status
- Open, Classified
- Product code
- OWB
- Quantity affected
- 4 units
- Distribution
- Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Bangladesh, Belgium, Brazil, Cayman Islands, Chile, China, Colombia, Costa Rica, Cuba, Czech Republic, Denmark, Domini
- Date initiated
- 2025-12-15
- Date posted
- 2026-01-13
- Location
- Best, Netherlands
Monitor recalls programmatically
Recalls and clearances share one API, with change alerts when new recalls post.
Related 510(k) clearances
| Record | Firm | Date |
|---|---|---|
| Allura Xper FD series and Allura Xper OR Table series (K161563) | Philips Medical Systems Nederland B.V. | 2016-07-29 |
| ALLURA XPER FD SERIES / ALLURA XPER OR TABLE SERIES (K141979) | Philips Medical Systems Netherland BV | 2014-08-19 |
| ALLURA XPER FD; OR TABLE (K133292) | Philips Medical Systems Nederland B.V. | 2014-03-05 |
| ALLURACLARITY XPER FD SERIES X-RAY SYSTEM (K130638) | Philips Medical Systems Nederland B.V. | 2013-06-28 |
| ALLURA XPER FD OR TABLE SERIES; ALLURA XPER FD10 SERIES; ALLURA XPER FD20 SERIES; ALLURA XPER FD BIPLANE SERIES (K130842) | Philips Medical Systems Nederland B.V. | 2013-05-21 |
| ALLURA XPER OR TABLE SERIES (K102005) | Philips Medical Systems North America, Inc. | 2010-08-09 |