Allen Traction Boot, Modular, OT 1000 Traction Device Boot Assembly, Fracture Table Orthopedic Leg Holders, non-invasive

FDA device recall Z-0959-2013 · posted 2013-03-18 · Terminated

Recall details

Recalling firm
Allen Medical Systems, Inc.
Reason for recall
Some units may have been assembled incorrectly which could result in the boot breaking free from the mounting base and allowing the patient's leg to fall.
Action taken
Allen Medical Systems sent a "Urgent Medical Device Recall" letter dated February 28, 2013, to all affected customers. The letter identified the product the problem and the action needed to be taken. Customers were instructed to: 1. Immediately remove from service your Allen Traction Boot with the serial number(s) XXXXX 2. Contact Allen Medical Customer Service at (800) 433-5774 for an RMA number and instructions to return your Allen Traction Boot for inspection and repair if necessary. There will be no cost to you. 3. Return your Allen Traction Boot to Allen Medical as soon as possible. In order to minimize the inconvenience, we will repair your Allen Traction Boot and return the device to you within 48 hours of receipt.
Root cause
Process control
Status
Terminated
Product code
HST
Quantity affected
24 devices
Distribution
Worldwide Distribution- USA including the states of TX, NV, KY, GA, OH, PA, NY, AL, WA, IL and the countries of New Zealand and Japan.
Date initiated
2013-02-28
Date posted
2013-03-18
Date terminated
2013-08-07
Location
Acton, MA