All Serial Numbers/BD Pyxis CII Safe ES Tower Main, REF: 1116-00

FDA device recall Z-0371-2026 · posted 2025-10-27 · Open, Classified

Recall details

Recalling firm
CareFusion 303, Inc.
Reason for recall
Potential fluid ingress of anesthesia station or med station may result in smoke, system downtime and/or fire.
Action taken
On September 30, 2025, BD issue a Urgent Medical Device Recall Expansion notification to affected consignees via FedEx. BD is asking consignees to take the following actions: 1. Affected products can continue to be used per existing Manuals and Guides. 2. In the event of an accidental fluid spill on the device or in a drawer, users should following instructions outlined within the customer notification. 3. Review the customer letter attachment which provides warnings, cautions and recommendations regarding fluid ingress. 4. Complete the attached Customer Response Form and return to the BD contact noted on the form whether or not you have any of the impacted material.
Root cause
Under Investigation by firm
Status
Open, Classified
Product code
BRY
Quantity affected
443 units
Distribution
Worldwide distribution - US Nationwide and the countries of Canada, United Arab Emirates, Australia.
Date initiated
2025-09-30
Date posted
2025-10-27
Location
San Diego, CA