Atlas FDA

Alaris SE Pump (formerly Signature Edition Infusion Pump); Model Numbers: 7000, 7100, 7101, 7130, 7131, 7132, 7200, 7201

FDA device recall Z-1484-06 · posted 2006-09-19 · Terminated

Recall details

Recalling firm
Cardinal Health 303 Inc DBA Alaris Products
Reason for recall
This recall was initiated because of a potential for over infusion with all models of the Alaris SE Pumps (formerly the Signature Edition Infusion Pumps) caused by key bounce.
Action taken
Corrective action for this recall will include notifications (five letters) being sent (Return Receipt) to all customers with affected product alerting them to this action. The letters are addressed to the Director of Biomedical Engineering, Director, Materials Management, Hospital Administration, Purchasing, and the fifth letter to the Director of Nursing. The August 15, 2006, recall letters and the warning labels will be sent return receipt requested. The verification of receipt of the notifications will be tracked and return receipt exceptions will be followed up. Follow up activities will be documented including customer contact.
Root cause
Software design
Status
Terminated
Product code
FRN
Quantity affected
147, 461
Distribution
Nationwide and American Samoa, Guam, and Puerto Rico. Worldwide.
Date initiated
2006-08-10
Date posted
2006-09-19
Date terminated
2010-07-06
Location
San Diego, CA

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
SE INFUSION SYSTEM WITH MMS (K043590)Cardinal Health, Alaris Products2005-03-25
SIGNATURE EDITION INFUSION PUMP, SIGNATURE EDITION ADMINISTRATION SETS, MODELS 70XX, 71XX, 72XX, 7XXXX SERIES (K032147)Alaris Medical Systems, Inc.2003-07-25
ALARIS MEDICAL INFUSION SYSTEM ADMINISTRATION SETS (K022209)Alaris Medical Systems, Inc.2002-08-06
GEMINI PC-1,PC-2, PC-2TX,PC-4, SIGNATURE EDITION INFUSION PUMP (SE), MEDSYSTEM III INFUSION (MSIII), MEDSYSTEM PATIENT C (K012383)Alaris Medical Systems, Inc.2001-10-02
MODEL EZ-1 SING CHAN/EZ-2 DUAL CHANNEL INFUS PUMP (K931549)Y1993-10-12