Atlas FDA

AirLife Nebulizer Heater, Catalog Code 2M8021 The AirLife Nebulizer Heater is part of the heated nebulizer system intend

FDA device recall Z-3016-2011 · posted 2011-08-12 · Terminated

Recall details

Recalling firm
Carefusion 211 Inc
Reason for recall
The recall was initiated because CareFusion's Investigation has confirmed the Nebulizer Heater can, over time, develop a failure mode that results in an internal electrical short. The short can result in smoke and sparks being emitted from the device. In the presence of oxygen enriched environments and a fuel source, this failure represents a potential risk of flame. CareFusion is voluntarily rec
Action taken
Carefusion sent a URGENT PRODUCT RECALL letter dated July 15, 2011, to all affected customers. The letter identified the product, the problem, and the action needed to be taken by the customer. Customers were instructed to examine their inventory for the affected product as indicated on the sample shown in the enclosed Attachment A-1 and remove any affected heaters present in their facility. Confirm the total number of affected units in their possession and enter the quantity and respective serial number on the enclosed Customer Response Card (Attachment A-2). Customers were instructed to contact CareFusion's Technical Support Department at 800-554-8933, (6:30 am to 5:00 pm PDST), to obtain an RMA (returned materials authorization) number for the return of the affected product. Upon receipt of their response, CareFusion will ship to their customers new replacement Nebulizer Heaters in exchange for the affected units. Customers were instructed to acknowledge receipt of this communication by completing the enclosed Customer Response Card and faxing or sending a copy by return mail to the address indicated on the Customer Response Card, (Attachment A-2).
Root cause
Other
Status
Terminated
Product code
CAF
Quantity affected
22,670 units total
Distribution
Nationwide Distribution
Date initiated
2011-07-13
Date posted
2011-08-12
Date terminated
2012-04-30
Location
Yorba Linda, CA

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
NEBULIZER HEATER (K780456)Travenol Laboratories, S.A.1978-04-12