AGI-2238 Portable 4 Cylinder Manifold Cart, for connecting medical gas cylinders.
FDA device recall Z-0038-2010 · posted 2009-10-14 · Terminated
Recall details
- Recalling firm
- Airgas Puritan Medical
- Reason for recall
- Yoke may become displaced. Yoke index safety system could become compromised, allowing a user to attach the wrong gas type to the yoke connection.
- Action taken
- Airgas issued a Medical Device Removal letter dated 8/6/09 to its customers informing them of the problem and the need to have a yoke back-up plate installed on the Manifold Block Assembly.
- Root cause
- Component design/selection
- Status
- Terminated
- Product code
- CAN
- Quantity affected
- 208
- Distribution
- Nationwide distribution.
- Date initiated
- 2009-08-06
- Date posted
- 2009-10-14
- Date terminated
- 2010-01-26
- Location
- Radnor, PA