ADVIA Centaur Intact PTH (iPTH) Assay -assay (lot 138) Cat. Nos. 100 Test kit 04385983 (129501) 500 Test kit 06587575 (1
FDA device recall Z-1186-2012 · posted 2012-03-07 · Terminated
Recall details
- Recalling firm
- Siemens Healthcare Diagnostics, Inc
- Reason for recall
- Poor sensitivity and imprecision on ADVIA Centaur systems; Serum samples low bias; EDTA plasma samples low bias
- Action taken
- Siemens sent an "URGENT FIELD SAFETY NOTICE" dated October 2008 to all affected customers. The letter identifies the product, problem, and actions to be taken by the customers. The letter instructed customer to immediately discontinue the use of ADVIA Centaur Intact PTH (iPTH) assay kits that contain intact PTH (iPTH) reagent lot 138. It also recommended that for serum samples analyzed on the ADVIA Centaur CP system, the content of the Urgent Field Safety Notice be discussed with their laboratory director to determine the need to review previous test results, conduct patient follow up, and repeat testing for iPTH results that were less than the upper limit of the reference range, and had a companion calcium result that was elevated or at the upper limit of normal. Additionally, a Completion Notification form was attached for customers to complete and return. Customers should contact their local technical support provider for questions regarding this recall.
- Root cause
- Process control
- Status
- Terminated
- Product code
- CEW
- Quantity affected
- 100 Test =5768; 500 Test =1778
- Distribution
- Worldwide Distribution-USA (nationwide) and the countries of Canada, Argentina, Chile, Columbia, Egypt, Finland, France, Hong Kong, India, Israel, Japan, Malaysia, Mexico, Netherlands, New Zealand, Po
- Date initiated
- 2008-10-03
- Date posted
- 2012-03-07
- Date terminated
- 2012-03-08
- Location
- East Walpole, MA