ADVIA Centaur CA19-9 Assay Cat. Nos. 123521 (03481938) - 250 Test kit w/o Calibrators 123519 (04612750) - 50 Test kit w/
FDA device recall Z-0489-2012 · posted 2012-01-11 · Terminated
Recall details
- Recalling firm
- Siemens Healthcare Diagnostics, Inc
- Reason for recall
- Positive bias of 14% and higher for some individual points for CA 19-9 results generated on the ADVIA Centaur CP system when they are compared to CA 19-9 assay results generated on either the ADVIA Centaur or ADVIA Centaur XP systems
- Action taken
- Siemens Healthcare Diagnostics issued a Urgent Device Notification email dated November 11, 2011, to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. The letter instructed the customers conduct a look back if they are using the ADVIA Centaur CA 19-9 interchangeably with the ADVIA Centaur or ADVIA Centaur XP systems and the ADVIA Centaur CP system and to review the contents of the recall notice with their Laboratory Director. A confirmation fax-back form is included to ensure that all customers worldwide have been notified and have complied with the mandatory action For further questions please call (508) 359.3825.
- Root cause
- Nonconforming Material/Component
- Status
- Terminated
- Product code
- NIG
- Quantity affected
- 50 Test Kits =21,160; 250 Test Kits = 21,593
- Distribution
- Worldwide Distribution -- USA (nationwide) and the countries of Australia, Malyasia, Korea, Hong Kong, Singapore, Myanmar, Mexico, Egypt, Taiwan, South Africa, Brazil ,India, Japan, Argentina, Chile,
- Date initiated
- 2011-11-11
- Date posted
- 2012-01-11
- Date terminated
- 2013-04-12
- Location
- East Walpole, MA