Atlas FDA

AdvanDx GNR Traffic Light PNA FISH, In Vitro Diagnostic for Culture Identification Kit for Escherichia coli,and/or Klebs

FDA device recall Z-0112-2013 · posted 2012-10-22 · Terminated

Recall details

Recalling firm
Advandx Inc
Reason for recall
GNR Traffic Light PNA Fish may have false positive results with Pseudomonas putida and Pseudomonas fulva
Action taken
AdvanDx sent a "GNR Traffic Light PNA FISH (KT011) Package Insert Replacement" letter dated September 14, 2012 to all affected customers. The letter identified the product, problem, and actions to be taken by the customers. All customers received an Acknowledgement Form to complete and return to AdvanDx so that the effectiveness of the replacement can be evaluated. Contact AdvanDx Technical Support at 1-866-376-0009 for questions regarding this notice.
Root cause
Device Design
Status
Terminated
Product code
JSS
Quantity affected
187 kits
Distribution
Worldwide Distribution-USA (nationwide) and the countries of Austria, Denmark, Italy, Germany, Finland, and UK.
Date initiated
2012-09-14
Date posted
2012-10-22
Date terminated
2013-03-05
Location
Woburn, MA