AD 7 Patient table supplied with Allura Xper FD10 and Allura Xper FD10/10. Product is non-tilting but can be purchased w
FDA device recall Z-1642-2008 · posted 2008-09-01 · Terminated
Recall details
- Recalling firm
- Philips Medical Systems North America Co. Phillips
- Reason for recall
- Potential for radiologic patient table to become immobile and unable to move again due to force sensor sensitivity to electromagnetic radiation.
- Action taken
- On 03/22/2004, firm sent letters via certified mail to notify customers of issue. Letter advises customers of potential risk for patients and users, and states to not use system for critical examinations or treatments if table movements suddenly stops or makes a freeing movement. The letter indicates a representative will contact customer for installation of a Hardware filter, free of charge, to prevent Electromagnetic interference (EMI) from affecting the force sensor. Please contact Philips at 1-800-722-9377 #5, #4, #1.
- Root cause
- Device Design
- Status
- Terminated
- Product code
- IXI
- Quantity affected
- 39 units
- Distribution
- Product is distributed to hospitals and medical centers nationwide to the following states: ND, FL, KY, MN, VA, ME, PA, NJ, CT, TX, CA, OH, TN, OR, GA, NY, SC, MO, WA, IL.
- Date initiated
- 2008-03-14
- Date posted
- 2008-09-01
- Date terminated
- 2012-07-18
- Location
- Bothell, WA