Acuson CV70 Ultrasound systems Ultrasonic Pulsed echo/doppler imaging system with Software revisions less than 4.0.0b, m

FDA device recall Z-2212-2008 · posted 2008-09-19 · Terminated

Recall details

Recalling firm
Siemens Medical Solutions USA, Inc.
Reason for recall
System may either display incorrect mechanical and thermal index values, or fail to display them altogether.
Action taken
Recall initiated on 06/04/2008. Under the Siemens update program, all consignees will receive a visit from Siemens service personal to install the revised software version. Siemens staff will coordinate with foreign distributors to arrange for service abroad.
Root cause
Software design
Status
Terminated
Product code
IYN
Quantity affected
1230 units
Distribution
Worldwide Distribution.
Date initiated
2008-06-04
Date posted
2008-09-19
Date terminated
2010-12-20
Location
Mountain View, CA