Accuracy Cyberknife Robotic Radiosurgery System, medical charged particle radiation therapy device. Model number 020700

FDA device recall Z-1862-2008 · posted 2008-09-16 · Terminated

Recall details

Recalling firm
Accuray Inc
Reason for recall
Couch may move unexpectedly, which may result in patient impacting the linear accelerator.
Action taken
All affected sites were notified via customer advisory notice. The Standard Treatment couch firmware will be updated and all affected sites will be retrofitted. Contact Customer Service at 1-877-668-8667 (USA) or 1-408-716-4700 (non-USA) or customersupport@accuray.com for additional information.
Root cause
Software design
Status
Terminated
Product code
IYE
Quantity affected
105 units
Distribution
Product was distributed throughout the US and to China, Taiwan, Korea, Italy, Malaysia, Germany, Netherlands, Vietnam, Greece, Spain, and Puerto Rico.
Date initiated
2008-05-01
Date posted
2008-09-16
Date terminated
2010-11-30
Location
Sunnyvale, CA