Accumetrics part number 29001 VerifyNow Instrument, printe power supply
FDA device recall Z-0102-06 · posted 2005-10-28 · Terminated
Recall details
- Recalling firm
- Accumetrics Inc
- Reason for recall
- In 2001, separate reports of mild shock, sparking and smoke from Accumetrics PN 29001 power supply.
- Action taken
- Letters and replacement power supplies were sent to consignees by Fedex, 2-day with proof of delivery. End-user customers were instructed to return the affected power supply to Accumetrics, using the provided pre-paid shipping carton. Consignees who do not return the power supply within two weeks of the notification will be contacted by phone to assess compliance with the recall and to determine the status of the product at the site.
- Root cause
- Other
- Status
- Terminated
- Product code
- JOZ
- Quantity affected
- 11
- Distribution
- CA, FL, IL, LA, NH, NV, NY, OK, WA
- Date initiated
- 2005-09-13
- Date posted
- 2005-10-28
- Date terminated
- 2007-03-03
- Location
- San Diego, CA