Accumetrics part number 29001 VerifyNow Instrument, printe power supply

FDA device recall Z-0102-06 · posted 2005-10-28 · Terminated

Recall details

Recalling firm
Accumetrics Inc
Reason for recall
In 2001, separate reports of mild shock, sparking and smoke from Accumetrics PN 29001 power supply.
Action taken
Letters and replacement power supplies were sent to consignees by Fedex, 2-day with proof of delivery. End-user customers were instructed to return the affected power supply to Accumetrics, using the provided pre-paid shipping carton. Consignees who do not return the power supply within two weeks of the notification will be contacted by phone to assess compliance with the recall and to determine the status of the product at the site.
Root cause
Other
Status
Terminated
Product code
JOZ
Quantity affected
11
Distribution
CA, FL, IL, LA, NH, NV, NY, OK, WA
Date initiated
2005-09-13
Date posted
2005-10-28
Date terminated
2007-03-03
Location
San Diego, CA