Accu-Chek Inform II Base Unit as a part of the Accu-Chek Inform II Blood Glucose Monitoring System Accu-Chek inform II B
FDA device recall Z-1172-2016 · posted 2016-03-14 · Terminated
Recall details
- Recalling firm
- Roche Diagnostics Operations, Inc.
- Reason for recall
- Accu-Chek Inform II Base Unit might produce physical transmission errors in the form of data loss in the communication between the meter and the Data Management Systems (DMS). The issue can lead to the data loss or in the worst case to an erroneous assignment of the patient data (patient mismatch). The issue will only occur at sites using POTCT1-A communication via USB.
- Action taken
- Roche sent an URGENT MEDICAL DEVICE RECALL CORRECTION letter dated February 11, 2016, via UPS Ground (receipt required) to the consignee. The Consignee instructions were as follows: If currently using USB connectivity, follow the instructions outlined in this Urgent Medical Device Correction (UMDC) to determine if your base unit is affected by the issue. Until the new software is available, do not use USB connectivity if the base unit is affected by this issue; switch the connectivity method to Ethernet (TCP/IP) or Wi-Fi. Roche is not requesting return of affected product. Any returned product will be managed according to current procedures. Consignees with questions were instructed to contact the ACCU-CHEK Customer Care Service Center at 1-800-440-3638. For questions regarding this recall call 317-521-3911.
- Root cause
- Software Design Change
- Status
- Terminated
- Product code
- LFR
- Quantity affected
- US 5,604 devices, OUS 91,925
- Distribution
- Nationwide Distribution including AK, AL, AR, AZ, CA, CO, CT, DC, FL, GA, IA, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, SC, TN, TX, UT, VA, WA, WI
- Date initiated
- 2016-02-11
- Date posted
- 2016-03-14
- Date terminated
- 2017-05-30
- Location
- Indianapolis, IN