Accu-Chek Comfort Curve glucose test strips; part numbers 2030420, 2030365, 2030373, 2030381, 3000133 and 3000141.
FDA device recall Z-0715-03 · posted 2003-04-09 · Terminated
Recall details
- Recalling firm
- Roche Diagnostics Corp.
- Reason for recall
- Crack in the bottom of the test strip vial will cause erroneously low or high blood glucose readings.
- Action taken
- Urgent Product Correction letters dated 2/12/03 were issued to pharmacists, specific health care providers, and the laboratory manager at customer accounts. Letter recipients were asked to advise their patients to inspect their vial for a crack prior to each use.
- Root cause
- Other
- Status
- Terminated
- Product code
- LFR
- Distribution
- United States, Argentina, Australia, Brazil, Canada, Chile, China, Colombia, Denmark, Finland, France, Germany, Guatemala, Honduras, Hong Kong, Indonesia, Japan, Malaysia, Mexico, New Zealand, Norway,
- Date initiated
- 2003-02-12
- Date posted
- 2003-04-09
- Date terminated
- 2003-09-15
- Location
- Indianapolis, IN