Accu-Chek Advantage Plus blood glucose test strips; Reference number 04735102001 (All sold outside of the United States)

FDA device recall Z-0569-06 · posted 2006-02-28 · Terminated

Recall details

Recalling firm
Roche Diagnostics Corp.
Reason for recall
There may be a thin piece of plastic across the mouth of the vial, preventing the cap from sitting snugly onto the vial and, if present, this situation will cause erroneously low or high blood glucose readings.
Action taken
Notifications (letters) dated 2/10/06 were sent to healthcare professionals, retail pharmacies, hospitals and mail order firms asking them to provide their customers with information instructing the users to inspect their strip vials and to not use strips from defective vials. Mail order accounts were instructed to forward the recall notification to end users.
Root cause
Other
Status
Terminated
Product code
LFR
Distribution
Nationwide, Argentina, Australia, Brazil, Canada, Chile, China, Colombia, Ecuador, England, France, Germany, Hong Kong, India, Indonesia, Japan, Malaysia, Mexico, New Zealand, Pakistan, Panama, Peru,
Date initiated
2006-02-10
Date posted
2006-02-28
Date terminated
2007-02-08
Location
Indianapolis, IN