Access Vitamin B12 Reagent, Catalog Number 33000, Lot Number 439850

FDA device recall Z-2115-2025 · posted 2025-07-11 · Completed

Recall details

Recalling firm
Beckman Coulter, Inc.
Reason for recall
Beckman Coulter has received complaints of erroneous Vitamin B12 results when using Access Vitamin B12 Reagent lot 439850. The affected reagent lot may cause erroneously low or high Vitamin B12 results, leading to a delayed or improper diagnosis and/or repeat testing.
Action taken
An URGENT MEDICAL DEVICE RECALL notification letter dated 2/19/25 was sent to customers. ACTION: Discontinue using Access Vitamin B12 reagent lot 439850. Discard all remaining reagent packs from this lot following your laboratory procedures. Beckman Coulter recommends sharing the content of this letter with your laboratory and/or medical director regarding the need to review previous patient results. Contact your local Beckman Coulter representative for replacement product requests. Please share this information with your laboratory staff and retain this notification as part of your laboratory Quality System documentation. If you have forwarded any of the affected product listed above to another laboratory, please provide them a copy of this letter. So that we are assured you have received this important communication, please respond within 10 days in one of the following ways: Electronically, if you received this communication via email. Manually, complete and return the enclosed Response Form. If you have any questions regarding this notice, please contact our Customer Support Center: From our website: http://www.beckmancoulter.com You may request replacement product by contacting your local Beckman Coulter Representative for replacement.
Root cause
Process control
Status
Completed
Product code
CDD
Quantity affected
10,000 units
Distribution
International distribution to the country of Andorra, Bulgaria, Denmark, Estonia, France, Germany, Greece, Ireland, Italy, Maldives, Netherlands, Portugal, Romania, Spain, Switzerland, Turkey, United
Date initiated
2025-02-27
Date posted
2025-07-11
Location
Chaska, MN