Access PCT Reagent Pack, Catalog Number C53987, when used on Access 2 and UniCel Dxl Immunoassay Systems

FDA device recall Z-1112-2025 · posted 2025-02-05 · Open, Classified

Recall details

Recalling firm
Beckman Coulter Inc.
Reason for recall
Identified reagent lots are experiencing a high rate of calibration failures with failure code of "Max Iterate" on Access 2 and UniCel Dxl Immunoassay Systems. Reporting of results may be delayed due to the time required to achieve a passing calibration curve.
Action taken
An Urgent Medical Device Recall letter was sent during the week of December 30, 2024, to the affected customers via first class mail and email. If a passing calibration curve can be generated using the Access PCT reagent lots listed above, no action is needed. If repeated calibration failures with a failure code of Max Iterate are obtained using the Access PCT reagent lots listed above, discontinue using the affected reagent lot number.
Root cause
Under Investigation by firm
Status
Open, Classified
Product code
PTF
Quantity affected
6,994 US; 10,067 OUS
Distribution
Worldwide distribution - US Nationwide and the countries of Austria, Bahrain, Bangladesh, Belgium, Bosnia and Herzegovina, Brazil, Canada, Chile, China, Croatia, Czechia, Denmark, France, Germany, Gre
Date initiated
2025-01-02
Date posted
2025-02-05
Location
Brea, CA