Atlas FDA

Access Immunoassay Systems Total T4 Calibrators, Part Number: 33805. The Access Total T4 Calibrators are intended to cal

FDA device recall Z-0613-2013 · posted 2012-12-22 · Terminated

Recall details

Recalling firm
Beckman Coulter Inc.
Reason for recall
A recall communication was initiated because Beckman has identified that Access Total T4 calibrator lots 019818, 021654, and 024072 do not meet their 12 month expiration date claim.
Action taken
A recall communication was initiated on July 27, 2011 with Beckman forwarding a Product Corrective Action (PCA) letter with attached PCA response form to all their customers who purchased the Access Immunoassay Systems Total T4 Calibrators. The letter provides the customers with an explanation of the problems identified and actions to be taken. Customers were instructed to complete and return the enclosed response form within 10 days. Customers can call for product replacement in the US, by contacting Client Services at (800) 526-3821, option 1. Customers in Canada can contact Customer service at (800) 463-7828. Customers with questions regarding this notice can contact Customer Support Center at http:www.beckmancoulter.com or call (800) 854-3633 in the US and Canada.
Root cause
Nonconforming Material/Component
Status
Terminated
Product code
JIS
Quantity affected
6,634 units total (2,742 units in US)
Distribution
Worldwide Distribution-USA (nationwide) and the countries of Argentina, Australia, Austria, Bangladesh, Belgium, Brazil, Canada, Chile, China, Colombia, Costa Rica, Czech Republic, Egypt, El Salvador,
Date initiated
2011-07-27
Date posted
2012-12-22
Date terminated
2012-12-26
Location
Brea, CA