Access Immunoassay Systems Total hCG. The Access Total hCG assay is a paramagnetic particle, chemiluminescent immunoassa

FDA device recall Z-0050-2012 · posted 2011-10-17 · Terminated

Recall details

Recalling firm
Beckman Coulter Inc.
Reason for recall
The recall was initiated because Beckman Coulter has confirmed that certain Access Total hCG reagent packs show decreased stability from the date of manufacture, which may lead to increased imprecision.
Action taken
The recall communication was initiated on March 18, 2009 with Beckman Coulter forwarding an Urgent: Product Corrective Action (PCA) letter dated March 13, 2009 with attached PCA Response Form (via US Postal Service for US Customers) to all customers who purchased the Access Immunoassay Systems total hCG. The letter provides the customers with an explanation of the problem identified and an action to be taken. Customers were instructed to discontinue use of the reagent lot numbers 824542 and 825652, and discard any unused reagent packs. Customers were asked to review their historical records for low level quality control to ensure that the assay performance for the referenced lot numbers has been within their laboratory specifications. They were also asked to share the information with their laboratory staff and retain the letter for their Quality System documentation, and then complete and return the enclosed response form within 10 days so that they may be assured that consignees received the notification. Finally they were told to contact Customer Technical Support Center at 1-800-854-3633 in the US and Canada, or their local Beckman Coulter Representative with any questions.
Root cause
Nonconforming Material/Component
Status
Terminated
Product code
JHI
Quantity affected
14,764
Distribution
Worldwide Distribution -- US, Algeria, Angola, Argentina, Armenia, Australia, Bangladesh, Belgium, Brazil, Bulgaria, canada, Chile, China, Colombia, Czech Republic, Egypt, El Salvador, Ethiopia, Finla
Date initiated
2009-03-05
Date posted
2011-10-17
Date terminated
2012-04-27
Location
Brea, CA