FDA device recall Z-1318-2010 · posted 2010-04-12 · Terminated

Recall details

Recalling firm
Beckman Coulter Inc.
Reason for recall
The recall was initiated because Beckman Coulter has confirmed customer reports that different results have been obtained using the same patient samples on Access/Access 2 and UniCel Dxl platforms. Values obtained with UniCel Dxl systems have been demonstrated to have a positive bias compared to values obtained with Access or Access 2 systems.
Action taken
An "URGENT: PRODUCT CORRECTIVE ACTION (PCA)" letter dated February 5, 2010, was sent to the customers. The letter describes the product, problem and action to be taken by the customers. The customers are recommended to immediately discontinue using AccuTnI Reagents and Calibrators on the UniCel DxI system until further notice. If this is not possible, the customers are directed to notify their stakeholders of the bias until further notice. The customers should complete, fax (786-639-4000) or mail the enclosed RESPONSE FORM. If consignees need assistance or have any questions regarding the notification, they were instructed to contact Technical Support at 1-800-854-3633 in the United States and Canada.
Root cause
Other
Status
Terminated
Product code
MMI
Quantity affected
196,071 kits
Distribution
Worldwide distribution: USA and Canada
Date initiated
2010-02-06
Date posted
2010-04-12
Date terminated
2012-05-09
Location
Brea, CA

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
ACCUTNI AND ACCUTNI CALIBRATORS ON THE ACCESS IMMUNOASSAY SYSTEMS, MODEL #33340, 33345 (K021814)Beckman Coulter, Inc.2002-09-27
ACCUTNI AND ACCUTNI CALIBRATORS ON THE ACCESS IMMUNOASSAY SYSTEM # 33340, 33345 (K010429)Beckman Coulter, Inc.2001-06-05