Access Immunoassay system CEA, in vitro diagnostic, Part No. 33200

FDA device recall Z-1313-06 · posted 2006-08-01 · Terminated

Recall details

Recalling firm
Beckman Coulter Inc
Reason for recall
Firm has confirmed that the Access CEA Assay may generate decreased values with certain frozen platelet specimens.
Action taken
A Product Corrective Action letter was sent via US mail the week of April 10, 2006 to all Access CEA customers informing them that the Access CEA Assay could generate decreased values with certain frozen patient specimens. Although they have found it at a low occurrence, it is recommended that specimens should not be frozen prior to analysis until further notice. The letter informs customers that Beckman Coulter has verified sample stability through at least 14 days when properly stored at 2-8 degrees Celsius and shipping stability for 4 days at 2-8 degrees Celsius using cool packs. Firm also advises customers to contact customer technical support or local Beckman Coulter Representative if laboratory is unable to avoid freezing specimens and to review results from frozen CEA samples to ensure that the match the total clinical presentation per the customer''s established laboratory procedure. The letter requests that the notification remain part of the laboratory Quality System documentation and that the response form be returned to the firm within 10 days.
Root cause
Other
Status
Terminated
Product code
DHX
Quantity affected
775
Distribution
Nationwide and Canada
Date initiated
2006-04-12
Date posted
2006-08-01
Date terminated
2012-04-03
Location
Brea, CA

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
ACCESS CEA REAGENTS ON THE ACCESS IMMUNOASSAY ANALYZER MODEL 33200, 33205, 33206 & 33209 (K981985)Beckman Coulter, Inc.1998-09-24