ABL800 FLEX Model Numbers 393-800 and 393-801 The ABL800 FLEX analyzers are intended for: " In Vitro testing of samples

FDA device recall Z-2426-2024 · posted 2024-07-24 · Open, Classified

Recall details

Recalling firm
Radiometer Medical ApS
Reason for recall
Issue with analyzer when the pH of the calibration solution decreases during the in-use period potentially because of bacterial growth in the calibration solution bottles. This may result in a probability of reporting biased out-of-specification pH results on blood samples.
Action taken
On 6/10/2024, the firm send an "URGENT: MEDICAL DEVICE RECALL" Letter to customer informing them that an internal technical study was carried out based on reports from internal and external users regarding pH sensitivity and pH drift errors for calibrations and measured QC pH values out of range. The study concludes that there is a possibility of reporting biased out-of-specification pH results on blood samples. This may occur if the calibration solution's pH value decreases during the in-use period due to bacterial growth in the calibration solution bottles CAL1 and/or CAL2. Customers are instructed to: -Immediately you implement either "Procedure 1" for customers who run daily quality controls or "Procedure 2" for customers who do not. The procedures are described on the subsequent pages of the Recall letter. This must be carried out regardless of which patient groups are measured on the analyzer. It will ensure that biased out-of-specification pH results that may lead to the risks above are not reported. PROCEDURE 1, for customers who run daily quality controls The individual steps of the procedures are described in the Operators Manual - Troubleshooting Messages (chapter 11). PROCEDURE 2, for customers who do not run daily quality controls The individual steps of the procedures are described in the Operators Manual - Troubleshooting Messages (chapter 11). For questions/assistance, contact the firm at 1-800-736-0600, Monday - Friday 9am to 7pm (EST).
Root cause
Nonconforming Material/Component
Status
Open, Classified
Product code
CHL
Quantity affected
27,127 analyzers
Distribution
Worldwide - U.S. Nationwide distribution including in the states of AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NY, OH, OK,
Date initiated
2024-05-31
Date posted
2024-07-24
Location
Bronshoj, Denmark