Atlas FDA

ABL77 Cal Pack, Model Number 944-069, Lot Number 17323, SenDx Medical Inc., A Radiometer Company

FDA device recall Z-1173-2007 · posted 2007-08-18 · Terminated

Recall details

Recalling firm
Sendx Medical Inc
Reason for recall
Error in the Ca12 calibration value for pCO2 recorded in the cal pack barcode for lot 17323. This error will result in falsely elevated pCO2 results at the high end of the measuring range.
Action taken
A total of four (4) distributors were sent the recall package (Field Action Notice) by email on 6-29-07. The distributors were provided background information and specific instruction that all Cal packs from lot 17323 should be removed from customer sites and destroyed. The Recall package included a Management Cover for Field Action Notes, a Customer Letter and the Field Action Notice #915-261. All communications regarding field actions associated with this recall were distributed using the Radiometer Field Action Note (FAN) system.
Root cause
Other
Status
Terminated
Product code
CHL
Quantity affected
69
Distribution
Worldwide Distribution --- USA including state of OH and countries of China, Denmark Latvia, Mexico & Spain.
Date initiated
2007-06-19
Date posted
2007-08-18
Date terminated
2011-11-07
Location
Carlsbad, CA

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
ABL 77 PH, BLOOD GAS AND ELECTROLYTE ANALYSIS SYSTEM (K994346)Sendx Medical, Inc.2000-02-01