ABL77 Cal Pack, Model Number 944-069, Lot Number 17323, SenDx Medical Inc., A Radiometer Company
FDA device recall Z-1173-2007 · posted 2007-08-18 · Terminated
Recall details
- Recalling firm
- Sendx Medical Inc
- Reason for recall
- Error in the Ca12 calibration value for pCO2 recorded in the cal pack barcode for lot 17323. This error will result in falsely elevated pCO2 results at the high end of the measuring range.
- Action taken
- A total of four (4) distributors were sent the recall package (Field Action Notice) by email on 6-29-07. The distributors were provided background information and specific instruction that all Cal packs from lot 17323 should be removed from customer sites and destroyed. The Recall package included a Management Cover for Field Action Notes, a Customer Letter and the Field Action Notice #915-261. All communications regarding field actions associated with this recall were distributed using the Radiometer Field Action Note (FAN) system.
- Root cause
- Other
- Status
- Terminated
- Product code
- CHL
- Quantity affected
- 69
- Distribution
- Worldwide Distribution --- USA including state of OH and countries of China, Denmark Latvia, Mexico & Spain.
- Date initiated
- 2007-06-19
- Date posted
- 2007-08-18
- Date terminated
- 2011-11-07
- Location
- Carlsbad, CA