Abbott Diabetes Care Optium EZ Blood Glucose Test Strips;
FDA device recall Z-0995-2011 · posted 2011-02-09 · Terminated
Recall details
- Recalling firm
- Abbott Diabetes Care, Inc.
- Reason for recall
- Certain lots of Precision Xtra, Precision Xceed Pro, MediSense Optium, Optium, OptiumEZ and ReliOn Ultima Blood Glucose Test Strips may generate erroneously low blood glucose results.
- Action taken
- On 12/22/10, Abbott Diabetes Care notified customers via Press Release and letters explaining the reason for the recall and requesting the products be returned.
- Root cause
- Process control
- Status
- Terminated
- Product code
- NBW
- Quantity affected
- 359 total lots, 1,000,000 strips per lot.
- Distribution
- Products were released for distribution in US and Puerto Rico.
- Date initiated
- 2010-12-22
- Date posted
- 2011-02-09
- Date terminated
- 2011-10-25
- Location
- Alameda, CA