Abbott brand FreeStyle Blood Glucose Meter, distributed by Abbott Diabetes Care, 1420 Harbor Bay Pwky. Alameda, CA 94502

FDA device recall Z-1306-06 · posted 2006-08-03 · Terminated

Recall details

Recalling firm
Abbott Diabetes Care, Inc.
Reason for recall
The products may encounter display problem, 'Er 4' message, during prolonged use when the low battery symbol is displayed. The situation can render the meter either inoperable or operable with invalid user configuration data including selectable unit of measure, and strip calibration code.
Action taken
On 5/22/06, the firm issued letters via Federal Express to all its direct consignees, informing them of the affected products and providing instructions for the corrective actions.
Root cause
Other
Status
Terminated
Product code
NBW
Quantity affected
approx. 2,365,000 meters (both recall numbers)
Distribution
Worldwide
Date initiated
2006-05-22
Date posted
2006-08-03
Date terminated
2006-12-21
Location
Alameda, CA