Abaxis brand Piccolo Comprehensive Metabolic Panel; Used with the Piccolo Blood Chemistry Analyzer or Piccolo xpress Che

FDA device recall Z-1066-2011 · posted 2011-02-09 · Terminated

Recall details

Recalling firm
Abaxis, Incorporated
Reason for recall
Abaxis has determined that during the manufacture of product lot 0396BC5 may report inaccurate values of Alanine Aminotransferase (ALT) when used in the Piccolo Blood Chemistry Analyzer or Piccolo xpress Chemistry Analyzer.
Action taken
The firm, Abaxis Inc., sent an "URGENT PRODUCT RECALL NOTICE" dated November 19, 2010, to all customers. The notice described the product, problem and action to be taken by the customers. The customers were instructed to check their inventory of the lot for boxes #0665 or higher; if found, to discontinue use; contact Abaxis for immediate replacement; complete and return the enclosed PRODUCT RECALL RESPONSE form to Abaxis Technical Support via fax at 1-877-349-2087, and ensure that their laboratory staff is informed of this recall notice. If you have any questions regarding this Product Recall Notice, please contact Abaxis Medical Technical Support at 800-822-2947, or at medtechsupport@abaxis.com.
Root cause
Other
Status
Terminated
Product code
CKA
Quantity affected
797 units
Distribution
Worldwide distribution: USA and countries of: Kowloon and Germany.
Date initiated
2010-11-19
Date posted
2011-02-09
Date terminated
2011-08-23
Location
Union City, CA