ABACUS TPN (Total Parenteral Nutrition) Calculation software, Product codes: 8300-0167 (Abacus V3.1 CE), 8300-0168 (Abac
FDA device recall Z-1376-2022 · posted 2022-07-22 · Terminated
Recall details
- Recalling firm
- Baxter Healthcare Corporation
- Reason for recall
- There is a potential risk of medication error when using specific Abacus Order Entry & Calculation software.
- Action taken
- Baxter sent an Urgent Medical Device Correction notice on 06/22/2022 by letter. The notice explained the problem and the hazard and recommended the following: customers refrain from making updates to bag label templates and to contact Baxter Technical Support if updates to bag label templates are needed. Please continue to follow the instructions in the Abacus User and Abacus Configuration Guides when creating an order, and have a pharmacist thoroughly review all order outputs including the printed label for accuracy and safety. As mentioned above, please ensure your facility processes include pharmacy checks during the compounding process as well as a nursing check. Baxter will communicate to customers when the software upgrade becomes available. At that time additional instructions to obtain the software upgrade will be provided.
- Root cause
- Software design
- Status
- Terminated
- Product code
- NEP
- Quantity affected
- 9 units
- Distribution
- United States (including Puerto Rico), Argentina, Bermuda, Chile, Colombia, Dominican Republic, Panama, Canada, Vietnam, Hong Kong, Singapore, and EMEA
- Date initiated
- 2022-06-22
- Date posted
- 2022-07-22
- Date terminated
- 2024-10-09
- Location
- Deerfield, IL