9131-001 Defibrillation Electrodes, Lot No. 141125-02 Used in Powerheart G3 9300A, 9300E, 9300P, 9390A, and 9390E AEDs C
FDA device recall Z-0829-2016 · posted 2016-03-09 · Terminated
Recall details
- Recalling firm
- Cardiac Science Corporation
- Reason for recall
- Defibrillation electrodes may increase electrical impedance over time. If impedance becomes too high, the AED will fail the electrode self-test and it will not be Rescue Ready, the AED STATUS INDICATOR will be RED and the AED will beep. This is normal behavior when the impedance limit is exceeded. The high impedance electrodes require replacement.
- Action taken
- Cardiac Science sent an Urgent: Voluntary Medical Device Recall letter dated January 2016 starting on January 14, 2014. The letters were either addressed to Distributor or Customer. The letter described the problem and the product involved in the recall. Requested consignees to replace any electrodes from the lot 141125-02 and return them to Cardiac Science. Furthermore, it advised consignees to contact their customers or members of their facilities who need to be made aware by providing a copy of the notification letter and replacement electrodes if they were a distributor. For questions contact Cardiac Science Technical Support at 1-800-426-0337, option 1 in USA or +1-262-953-3500 OUS or via e-mail a9131cardiacscience.com. For questions regarding this recall call 800-426-0337.
- Root cause
- Nonconforming Material/Component
- Status
- Terminated
- Product code
- MKJ
- Quantity affected
- 10,943
- Distribution
- Worldwide Distribution - US all states including PR except ND & RI. International Distribution to Argentina, Australia, Belgium, Bermuda, Canada, Chile, Colombia, Ecuador, France, Germany, Hong Kong,
- Date initiated
- 2016-01-14
- Date posted
- 2016-03-09
- Date terminated
- 2016-12-21
- Location
- Waukesha, WI