8MM,MICRO BIPOLAR FORCEPS; Intended to assist in the accurate control of Intuitive Surgical Endoscopic Instruments.

FDA device recall Z-0431-2015 · posted 2014-11-25 · Terminated

Recall details

Recalling firm
Intuitive Surgical, Inc.
Reason for recall
Deviations in reprocessing steps from those stated in the reprocessing instructions can cause surface degradation of the housing and/or accelerate mechanical wear of the instrument.
Action taken
Urgent Medical Device Recall letters were sent to affected sites on November 3, 2014. Letters inform customers of the reason for recall, risk to health and the affected instrument. A table of all of the affected IS4000 instruments affected is attached to the letter indicating the part number and product name for each. Letters also inform customers that all instruments shipped since July 2014 are manufactured with a new material and not subject to recall. Customers are to take the following actions: 1. Ensure that all affected personnel are aware of the information. 2. Identify affected product in their possession 3. Contact ISI rep for replacement. 4. Complete and return the attached Acknowledgement From. Until replacement instruments are received, the old instruments may be used provided they do not show signs of degradation and that the reprocessing instructions are followed. Questions to 800-876-1310 Option 3 ( 6 am to 5 pm PST) or ics@intusurg.com.
Root cause
Labeling design
Status
Terminated
Product code
NAY
Quantity affected
509 Units, total all affected units
Distribution
US Distribution -- FL, TX, NY, NV, OR, IN, UT, and MA.
Date initiated
2014-11-03
Date posted
2014-11-25
Date terminated
2015-02-03
Location
Sunnyvale, CA

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
DA VINCI SURGICAL SYSTEM, ENDOWRIST INSTRUMENTS AND ACCESSORIES (K131861)Intuitive Surgical, Inc.2014-03-28