804117, Terumo Coronary Ostia cannula, 12 Fr, 10 (25 cm) long A cardiopulmonary bypass vascular catheter, cannula, or tu

FDA device recall Z-2555-2011 · posted 2011-06-14 · Terminated

Recall details

Recalling firm
Terumo Cardiovascular Systems Corporation
Reason for recall
The presence of adhesive and/or skiving of plastic material observed inside the distal tip of the cannula, which could lead to particulate and/or decrease in the internal diameter. The plastic skiving may detach and enter the patient, embolizing the coronary vasculature. The potential risk due to decrease in the internal tip diameter is inadequate delivery of cardioplegia and/or prolonged procedu
Action taken
Terumo Cardiovascular Systems Corporation sent an URGENT MEDICAL DEVICE REMOVAL notice dated May 16, 2011, to all affected customers. The notice identified the product, the problem, and the action to be taken. The letter described the problem, the potiential hazard, and the action to be taken by the consignee. Consignees were instructed to discontinue use of the product and return all product to Terumo. The letter provided a list of alternate products to be used instead of the recalled products. Consignees with questions should call 1-800-521-2818 Monday through Friday 8AM to 6PM.
Root cause
Process control
Status
Terminated
Product code
DWF
Quantity affected
4540
Distribution
Worldwide Distribution - USA (nationwide) and the countries of BELGIUM, BRAZIL, DOMINICAN REPUBLIC, UNITED ARAB EMIRATES (UAE), SINGAPORE, MALAYSIA, AND SOUTH AFRICA.
Date initiated
2011-05-16
Date posted
2011-06-14
Date terminated
2012-03-21
Location
Ann Arbor, MI