Atlas FDA

8000-021-002, Scopis ENT Software with TGS with Sofware Version: NOVA 3.6.0 RC16. Scopis is a next-generation solution f

FDA device recall Z-1885-2024 · posted 2024-05-23 · Open, Classified

Recall details

Recalling firm
Stryker Leibinger GmbH & Co. KG
Reason for recall
Unreleased software was installed on customer systems resulting in the visual feedback on the screen to show the points to be off from the physical reference point of the pointer or suction.
Action taken
On April 22, 2024 an URGENT MEDICAL DEVICE FIELD ACTION letters were sent to customers. Next Steps: 1. Locate any ENU. Quarantine and discontinue use until software on device has been confirmed. - Locate software version and serial number (see pg. 2 for instructions). - Your Stryker Sales Representatives may help, if needed. 2. Return the enclosed Business Reply Form (BRF) via email to instruments.recalls@stryker.com. - If Scopis ENT Software with TGS version NOVA 3.6.0 RC16 is installed, Stryker will arrange shipment of compliant software. 3. Once compliant software is received and installed, return the enclosed Software Update Form via email to instruments.recalls@stryker.com to confirm completion of the recall action steps. 4. Maintain this notification internally until all required actions have been completed. Please contact Lacey Marshall at 269-270-5950 or Instruments.recalls@stryker.com with questions or concerns. Report any serious adverse events or product quality problems to Stryker Instruments: 1-800-253-3210. Health care professionals and consumers may report serious adverse events or product quality problems with the use of this product to the FDA's MedWatch Adverse Event Reporting program either online, or by fax or phone. Online: ww.fda.gov/Safety/MedWatch/HowToReport/default.htm Fax: (800) FDA-0178 Phone: (800) FDA-1088.
Root cause
Software Manufacturing/Software Deployment
Status
Open, Classified
Product code
PGW
Quantity affected
705 units
Distribution
US Nationwide distribution.
Date initiated
2024-04-04
Date posted
2024-05-23
Location
Freiburg Im Breisgau, Germany

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
Scopis ENT Software, Scopis ENT Software with TGS, Scopis Planning Software, Electromagnetic Navigation Unit (K221098)Stryker Corporation2022-07-12