6-hole Pubic Symphysis plate (PN 70-0451). The pelvic bone plates for the pelvic ring are a series of plates with varyin
FDA device recall Z-1199-2017 · posted 2017-02-14 · Terminated
Recall details
- Recalling firm
- Acumed LLC
- Reason for recall
- The 6-hole Pubic Symphysis plate (PN 70-0451) is specified to be made of Commercially Pure Titanium (ASTM F67 - CP Grade 4). However, batch 383813 was manufactured out of Titanium Alloy (Ti-6Al- 4V ELI per ASTM F135 Ti alloy) and was distributed.
- Action taken
- Acumed sent an Urgent Medical Device Recall letter dated January 16, 2017, to all affected customers on January 19, 2017. Consignees were instructed to follow special instructions and send survey to RecallNotification@acumed.net. Domestic customers with questions or need to return products to Acumed, please contact Acumed Agent Inventory via email AgentInventory@acumed.net to obtain a Return Material Authorization (RMA) number or contact your agent directly. International customers with questions or need to return products to Acumed, please contact Acumed Business Service via email at BusinessServices@acumed.net or via phone 877-627-9957, choose Option 4 (US and Canada), or call 001-503-627-9957, choose option 4 (outside of US and Canada) have to obtain a Return Material Authorization (RMA) number.
- Root cause
- Mixed-up of materials/components
- Status
- Terminated
- Product code
- HRS
- Quantity affected
- 15 units
- Distribution
- Worldwide Distribution - Costa Rica, Malaysia, Mexico, South Africa, and Spain
- Date initiated
- 2017-01-19
- Date posted
- 2017-02-14
- Date terminated
- 2017-05-08
- Location
- Hillsboro, OR
Monitor recalls programmatically
Recalls and clearances share one API, with change alerts when new recalls post.
Related 510(k) clearances
| Record | Firm | Date |
|---|---|---|
| ACUMED PELVIC BONE PLATE SYSTEM (K122538) | Acumed, LLC | 2012-11-19 |