Atlas FDA

6-hole Pubic Symphysis plate (PN 70-0451). The pelvic bone plates for the pelvic ring are a series of plates with varyin

FDA device recall Z-1199-2017 · posted 2017-02-14 · Terminated

Recall details

Recalling firm
Acumed LLC
Reason for recall
The 6-hole Pubic Symphysis plate (PN 70-0451) is specified to be made of Commercially Pure Titanium (ASTM F67 - CP Grade 4). However, batch 383813 was manufactured out of Titanium Alloy (Ti-6Al- 4V ELI per ASTM F135 Ti alloy) and was distributed.
Action taken
Acumed sent an Urgent Medical Device Recall letter dated January 16, 2017, to all affected customers on January 19, 2017. Consignees were instructed to follow special instructions and send survey to RecallNotification@acumed.net. Domestic customers with questions or need to return products to Acumed, please contact Acumed Agent Inventory via email AgentInventory@acumed.net to obtain a Return Material Authorization (RMA) number or contact your agent directly. International customers with questions or need to return products to Acumed, please contact Acumed Business Service via email at BusinessServices@acumed.net or via phone 877-627-9957, choose Option 4 (US and Canada), or call 001-503-627-9957, choose option 4 (outside of US and Canada) have to obtain a Return Material Authorization (RMA) number.
Root cause
Mixed-up of materials/components
Status
Terminated
Product code
HRS
Quantity affected
15 units
Distribution
Worldwide Distribution - Costa Rica, Malaysia, Mexico, South Africa, and Spain
Date initiated
2017-01-19
Date posted
2017-02-14
Date terminated
2017-05-08
Location
Hillsboro, OR

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
ACUMED PELVIC BONE PLATE SYSTEM (K122538)Acumed, LLC2012-11-19