515B Pulse Oximeter, Refurbished - Dixtal owned product. Intended use is to provide continuous, non-invasive monitoring

FDA device recall Z-1306-2010 · posted 2010-04-08 · Terminated

Recall details

Recalling firm
Philips Healthcare Inc.
Reason for recall
Power cords' prongs may crack or fail -- Philips was notified that FDA is investigating whether certain types of Electri-Cord Manufacturing Company power cords used with medical devices may be defective in that the power cord's prongs may crack and fail at/or inside the plug. Potential for burns or electrical shock.
Action taken
Philips was notified that FDA is investigating whether certain types of Electri-Cord Manufacturing Company power cords used with medical devices may be defective in that the power cord's prongs may crack and fail at/or inside plug. The issue affects medical device power cords equipped with a plug that has a prong and ground-pin insert design and a black plastic bridge connecting the terminal prongs on the plug ("taller bridge" design). Medical device power cords that do not have a black bridge connecting the terminal prongs on the plug are not affected. Urgent Medical Device Correction letters were sent to Distributors describing the problem and giving them instructions for notifying customers and actions to be taken. Customers are to visually inspect the affected cords, complete the Customer Reply Power Cord Order Form, and follow the instructions to dispose of affected cords. Further information or support concerning this issue can be obtained by contacting a Philips representative at 1-800-722-9377.
Root cause
Nonconforming Material/Component
Status
Terminated
Product code
DQA
Quantity affected
48
Distribution
Worldwide Distribution -- UNITED STATES, ALGERIA, ARGENTINA, Aruba, AUSTRALIA, BAHRAIN, BANGLADESH, BELGIUM, BRAZIL, CAMBODIA, CANADA, CHILE, CHINA, COLOMBIA, COSTA RICA, CROATIA, CZECH REPUBLIC, Denm
Date initiated
2009-11-19
Date posted
2010-04-08
Date terminated
2016-12-08
Location
Andover, MA

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
NOVAMETRIX MDLS 500A, 505A, 515A, & 7000A OXIMETER (K893643)Novametrix Medical Systems, Inc.1989-10-26