45cm Laparoscopic Instruments Sterilization tray. Manufactured by: Stryker Endoscopy, 5900 Optical Court,San Jose, CA 95

FDA device recall Z-2612-2011 · posted 2011-06-22 · Terminated

Recall details

Recalling firm
Stryker Endoscopy
Reason for recall
During testing and review, Stryker Endoscopy discovered that the sterilization parameters for the 132 degree C gravity steam sterilization cycle in the IFU may not consistently sterilize all instruments in the 45cm Laparoscopic Instruments Tray.
Action taken
Stryker Endoscopy sent a "URGENT: DEVICE CORRECTION" letter dated June 8, 2011 to affected customers. The letter describes the product, problem, and actions to be taken by the customers. Customers were instructed to discard all old Instructions for Use sheets for their 45cm Laparoscopic Tray and download the new IFU from the URL link:http://stryker.com/45cm tray. Additionally, customers were instructed to complete and return an attached "acknowledgement of receipt" form and fax to 408-754-8378 or scan and e-mail to trays@stryker.com. Questions can be answered by calling 1-800-624-4422 or by e-mail at trays@stryker.com.
Root cause
Nonconforming Material/Component
Status
Terminated
Product code
KCT
Quantity affected
603
Distribution
Worldwide Distribution--USA (nationwide) and countries of Brazil, Canada, Chili, Germany , Korea, Mexico, Philippines, Poland, Portugal, Singapore, Sweden, Switzerland, and Taiwan.
Date initiated
2011-06-08
Date posted
2011-06-22
Date terminated
2011-10-25
Location
San Jose, CA