3M" Steri-Vac" Gas Sterilizer/Aerators, Model #'s 4XL,5XL, 8XL The 3M" Steri-Vac" Gas Sterilizer/Aerator is a compact un
FDA device recall Z-1318-2014 · posted 2014-04-02 · Terminated
Recall details
- Recalling firm
- 3M Company - Health Care Business
- Reason for recall
- Recently, 3M received a single report of a potential exposure to Ethylene Oxide (EO) at a hospital installation involving 3M" Steri-Vac" Gas Sterilizers. The cause was traced to a rare situation involving a failure of the EO cartridge puncture assembly. This assembly was only in production during the period November 10, 2006 through December 14, 2007 and in service during the period February 7, 20
- Action taken
- 3M sent an "Urgent Medical Device Field Correction" letter dated March 6, 2014. The letter was sent to all affected customers. The letter identified the product the problem and the action needed to be taken by the customer. You are being notified because our records indicate you received a unit that contains the cartridge puncture assembly that is the subject of this field correction. 3M will contact you to schedule an appointment for a field correction on any unit in the range of serial numbers provided in the table. 3M's goal is to service the affected units with minimal disruption to your facility. Units manufactured outside of the dates November 10, 2006 to December 14, 2007 are not affected. .For Further Information contact 3M Health Care Service Line at 1-800-228-3957 or your local 3M representative.
- Root cause
- Device Design
- Status
- Terminated
- Product code
- FLF
- Quantity affected
- 73 US, 346 OUS
- Distribution
- Worldwide Distribution - USA (nationwide) Distribution including the states of AL, AZ, CA,CO, FL, GA, HI, IL, IN, IA, KS, KY, MD, MA, MI, MN, MS, NE, NJ, NY, NC, ND, OH, OK,OA, SC, TX, WA, WV and WI.
- Date initiated
- 2014-03-17
- Date posted
- 2014-04-02
- Date terminated
- 2015-01-22
- Location
- Saint Paul, MN
Monitor recalls programmatically
Recalls and clearances share one API, with change alerts when new recalls post.
Related 510(k) clearances
| Record | Firm | Date |
|---|---|---|
| STERI-VAC GAS STERILIZER MODEL 8XL (K941748) | 3M Health Care, Ltd. | 1994-12-02 |
| STERI-VAC GAS STERILIZER MODEL 5XL (K902036) | 3M Company | 1990-07-19 |
| STERI-VAC #400C GAS STERILIZER (K812867) | 3M Company | 1981-11-02 |