Atlas FDA

3M Attest Super Rapid 5 Steam-Plus Steam Challenge Pack, REF 41482V; designed to qualify and monitor dynamic-air removal

FDA device recall Z-2639-2023 · posted 2023-09-26 · Open, Classified

Recall details

Recalling firm
3M Company - Health Care Business
Reason for recall
Specific lots of 3M Attest Super Rapid Readout Biological Indicators (1492V), 3M Attest Super Rapid 5 Steam-Plus Challenge (41482V, 41482VF), and 3M Attest Super Rapid Readout Steam Challenge (1496V) may contain biological indicators with a cap that melts, deforms and/or cracks following a steam sterilization cycle of 132C and 135C.
Action taken
3M issued an Urgent product recall notice to its consignees on 08/24/2023 by email and US mail. The notice explained the problem with the product and requested the following: Discontinue distribution of the product and dispose of the product. Distributors were directed to notify their customers. The firm is offering replacement with the return of the completed response form.
Root cause
Process control
Status
Open, Classified
Product code
FRC
Quantity affected
172,464 devices
Distribution
Worldwide distribution - US Nationwide and the countries of Argentina, Australia, Belgium, Canada, China, Colombia, Hong Kong, Italy, Japan, Peru, Poland, South Korea, Spain, Taiwan.
Date initiated
2023-08-22
Date posted
2023-09-26
Location
Saint Paul, MN

Monitor recalls programmatically

Recalls and clearances share one API, with change alerts when new recalls post.

Related 510(k) clearances

RecordFirmDate
3M Attest Super Rapid 5 Steam-Plus Challenge Pack (K193154)3M Company2020-02-07